Paediatric Interventional Cardiology Diagnostic Reference Levels in Ireland
Congress: ECR 2019
Poster No.: C-0907
Type: Scientific Exhibit
Keywords: Quality assurance, Radiation safety, Fluoroscopy, Radioprotection / Radiation dose, Paediatric, Cardiac
Authors: D. Murphy, S. McCoubrey; Dublin/IE
DOI:10.26044/ecr2019/C-0907

Results

Although all examinations contained numerous outliers, strong to moderate correlations were found between KAP and patient weight for the majority of procedures (see table 1).  The median (Q50) and 75th percentile (Q75) are presented for each examination as KAP per kg (see table 2).  An example of how these values translate to a variety of patient weights is illustrated in figure 1, for atrial septal defect closure. 

 

Other studies have reported diagnostic reference levels for all IC procedures taken together [4, 5].  Table 3 compares our IC data, taken together, with those reported in the literature.