Rapid bedside assessment of the renal function of patients undergoing contrast-enhanced CT. Is it a reliable approach for identifying patients at risk of a contrast medium adverse reaction ?. 
Congress: ECR 2013
Poster No.: C-2669
Type: Scientific Exhibit
Keywords: CT, Kidney, Contrast agents, Laboratory tests, Drugs / Reactions
Authors: A. Wibmer1, R. Nolz1, G. Heinz-Peer2; 1Wien/AT, 2St. Pölten/AT
DOI:10.1594/ecr2013/C-2669

Conclusion

The StatSensor® creatinine meter is a reliable tool in the screening for patients with renal dysfunction with high negative predictive values. Due to low positive predictive values, caution should be exercised in case of positive results, which should be validated by standard laboratory measurement.