ECR 2014 / C-2302
Correlation between Perfusion Micro-Bubble Ultrasound and Perfusion Contrast Enhanced MRI for Carotid Plaque Neovascularization
This poster is published under an open license. Please read the disclaimer for further details.
Congress: ECR 2014
Poster No.: C-2302
Type: Scientific Exhibit
Keywords: Tissue characterisation, Image registration, Segmentation, Contrast agent-intravenous, Ultrasound, MR, Vascular, Neuroradiology brain, Contrast agents
Authors: W. Abutaleb1, M. J. Graves2, J. H. Gillard2; 1Riyadh/SA; Cambridge/UK, 2Cambridge/UK
DOI:10.1594/ecr2014/C-2302

Results

  We have acquired interpretable images from 9 patients. The results show 8 out of the 9 cases correlate in the level of enahancement. The correlation measured between the (High/Low) avereged EI(plaque)/EI(lumen) dB-enhancement of microbubble values versus the mean of Ktrans of the plaques perfusion MRI. 

 

 

  The statistical values: Using Spearman Rank Correlation:

  Number of points = 9

  Spearman r = 0.6614 (corrected for ties)

 

 

   Test: Is r significantly different than zero? 

   The two-tailed P value is 0.0589, considered not quite significant.

   The P value may be inaccurate because of ties.


 

This table show the result for DCE-US and DCE-MR for each case:

 


Case number  DCE-MR Ktrans means: High=1, Low=0

DCE-US EI(plaque)/EI(lumen):

High=1, Low=0

1
1
1
2
1
1
3
1
1
4
0
1
5
1
1
6
1
1
7
1
1
8
0
0
9
1
1

Table 1: DCE-US vs. DCE-MR: values in two catogeries (High/Low)

 

Case number

 DCE-US

  EI(plaque)/EI(lumen)

DCE-MR

 Ktrans means

1
2.63E-01
0.068974
2
1.45E-01
0.06113525
3
3.68E-01
0.06950125
4
9.28E-02
0.04871425
5
3.89E-01
0.09110175
6
4.48E-02
0.077622
7
4.02E-02
0.07345025
8
3.31E-02
0.0491245
9
7.26E-02
0.087293

Table 2: DCE-US vs. DCE-MR the actual values

 

 

 The resulted Ktrans values (Fig. 7) ranged between 0.04 min^-1 and 0.09. We considered values equal or larger than 0.06 min^-1 as “High” values and lower values are “Low”. For the ultrsound (Fig.8) EI ratios, values ranged between 2.45e-02 and 3.89e-01. We considered values as “High” if higher than 4.00e-02.

  Noted in the results as well that the values range in MRI is narrower than that of Ultrasound. 

 